Cleaning The Gemcone Nebulizer - PerkinElmer AVIO 500 Customer Hardware And Service Manual

Icp optical emission
Table of Contents

Advertisement

Figure 5-32
Item
1
Cyclonic Spray Chamber
2
Tubing Assembly for the
GemCone Nebulizer (Part No.
N0770336)
3
Argon Gas Fitting

Cleaning the GemCone Nebulizer

With routine use it may become necessary to perform a thorough cleaning.
1. Remove the nebulizer from the end cap (loosen both thumbscrews).
2. Disconnect the argon gas fitting and the capillary tube connection from the
nebulizer.
3. Immerse the nebulizer in a solution of liquid laboratory cleaner (such as
"Microcleaning" solution, Cole-Parmer). Use the recommended concentration
of cleaning solution.
6
5
4
3
2
Connecting the GemCone nebulizer to the cyclonic spray chamber.
Description
GemCone Nebulizer
1
Item
Description
4
GemCone Nebulizer Assembly
5
Teflon Tubing-1 inch (Part No.
09985708)
6
Capillary Tubing (Part No. 09908265)
275