Removing and Replacing the Axial Flat Optic

You may need to remove and replace the Axial Flat Optic if you notice reduced
Axial sensitivity or counts.
Tools Needed:
1. Turn of the plasma.
If the plasma has been on, the torch glass will be very hot and can
!
cause serious burns. Wait five minutes after turning off the plasma
before you begin these maintenance procedures.
Warning
Avertissement: Si le plasma a été mis sur le verre de la torche sera
très chaude et peut causer des brûlures graves. Attendez cinq minutes
après avoir éteint le plasma avant de commencer ces procédures
d'entretien.
2. Slide the Axial Sink Cooling tube as far as you can to the right. See the
following figure.
Removing and Replacing the Axial Flat Optic
6mm Hex Mini Ratchet
249