Maintenance
268

Connecting the Nebulizer Argon Tubing (Cross-Flow End Cap, N0780546)

1. Pass the tubing (Part No. 09985723) through the nut (Part No. 09920545) and
attach the ferrule (Part No. 09920515) on the end of the tubing. See the
following figure.
2. Insert the ferrule with the tubing and nut to the Male Quick Disconnect (Part
No. N0775206). Connect the entire Quick Disconnect to the sample
introduction system. See the following figure.

Replacing the Nebulizer Argon Tubing (Cross-Flow End Cap, N0780546)

If you replace the argon tubing with new tubing, you will need to replace the
Swagelok nut as follows. See the previous figure.
1. Using a razor blade, make a straight cut on one end of the tubing. Wire cutters
or scissors do not make acceptable cuts.
2. Put the 1/8-in. Swagelok nut (Part No. 09903128), rear ferrule, front ferrule,
and tubing insert onto the tubing as shown in the figure below.
3. Finger-tighten the nut on the Brass Reducer Fitting.