Peristaltic Pump

The peristaltic pump requires routine maintenance to ensure that the system will
run properly. The pump tubing should be replaced when flat spots develop (on a
daily basis or after approximately eight hours of use). When replacing the pump
tubing, the drain tubing should be replaced at the same time since uneven wear can
cause the spray chamber to back up.

Replacement Pump Tubing

Replace the pump tubing when flat spots develop (after approximately eight hours
of use). When replacing the pump tubing, both the sample and drain tubing should
be replaced at the same time since uneven wear can cause the spray chamber to
back up.The following table gives part numbers for pump tubing for the sample and
drain.
Table 5-7.
Pump Tubing Solvent Type
Standard
Inorganic
Solvent Flex
Kerosene or xylene
Silicone
Organic solvents,
including MIBK, but not
kerosene or xylene
Viton
Kerosene or xylene, but
not MIBK.

Installing the Sample and Drain Tubing

The peristaltic pump is used to pump the sample into the nebulizer and pump the
waste out of the spray chamber. Tighten the adjustment screw (turn clockwise) up
to three turns until the bubbles are visible in the drain tubing. Then add three more
turns of the adjustment screws.
Pump Tubing
Inner Diameter
Sample: 0.76 mm (0.030 in.)
Drain: 1.14 mm (0.045 in.)
Sample: 0.76 mm (0.030 in.)
Drain: 1.14 mm (0.045 in.)
Sample: 0.76 mm (0.030 in.)
Drain: 1.14 mm (0.045 in.)
Sample: Orange-Orange
0.89 mm (0.035 in.)
Drain: Yellow-Yellow
1.40 mm (0.056 in.)
Peristaltic Pump
Part No.
09908587
09908585
00473550
09923037
00473552
N0691595
09923511
09923512
285