Connecting The System Components - PerkinElmer AVIO 500 Customer Hardware And Service Manual

Icp optical emission
Table of Contents

Advertisement

Installation
128
2. With the chiller running, top off the reservoir tank, reinstall the reservoir cap
and turn clockwise to tighten.
3. Install the reservoir access panel and secure it with its screws.

Connecting the System Components

!
Electrical Hazard
To prevent potential injury to yourself and damage to the instrument, switch
Warning
OFF all instruments in the system and disconnect them from the line power
supply before you perform the procedures described below.
Danger électrique
Pour éviter les risques de blessures pour vous et d'endommager l'appareil,
éteindre tous les instruments dans le système et les déconnecter de
l'alimentation de la ligne avant d'effectuer les procédures décrites ci-dessous.
The system component connections are as follows (see the following figure):
Extending Ethernet cables by making tandem connections will
Note
compromise instrument communication.
Make sure that the computer and printer share a common ground with the
Note
instrument
Ethernet cable from the computer to the instrument Ethernet connector on
the spectrometer side of the instrument. RS232 cable from an optional
autosampler to the computer.
The Avio 500 is equipped with an IEC 309 250 V 16/20A 2 pole plus
protective earth plug (PerkinElmer Part No. 09997530) that inserts into an
equivalent IEC 309 series receptacle (see Section 2, Electrical
Requirements).
Connect the computer and printer to electrical power as described in the
guides provided with these items.
.