Electrical Requirements - PerkinElmer AVIO 500 Customer Hardware And Service Manual

Icp optical emission
Table of Contents

Advertisement

Preparing Your Laboratory
68
A PolyScience Recirculating Chiller meets these requirements and is
recommended for the instrument. The PolyScience
available through PerkinElmer in the following two configurations:
230V, 60 Hz (Part No. N0772046)
240V, 50 Hz (Part No. N0772045)
For detailed information on PolyScience
requirements see the following section, Electrical Requirements.
The PolyScience
The headquarters for PolyScience Instruments is located at:
PolyScience Instruments, Inc.
6600 West Touhy Ave
Niles, Il 60714
USA: (800) 229-7569
Worldwide: (847) 647-0611
Fax: (847) 647-1155

Electrical Requirements

Avio 500
Grounding circuit continuity is vital for safe operation of this instrument.
Grounding is accomplished by use of an IEC60309 instrument power plug
furnished by the factory and not to be removed during or after installation.
Avertissement: Mise à la terre la continuité du circuit est essentiel pour une
utilisation sûre de cet instrument. La terre est effectuée par l'utilisation d'une
prise d'alimentation de l'instrument IEC60309 fourni par l'usine et ne doit pas
être retiré pendant ou après l'installation.
The Avio 500 ICP-OES is equipped with an IEC 309 250 V 16/20 A 2-pole
plus protective earth plug (Hubbell® Part No. C320P6SVL, PerkinElmer Part
No. 09997530) that inserts into a receptacle (Walther Part No. 410306,
PerkinElmer Part No. 09290304). Alternative surface mount receptacle
(Walther Part No. 111306, PerkinElmer Part No. 09290305). Both receptacles
are shipped with the instrument.
®
TM
WhisperCool
chiller comes with its own Instruction Manuals.
®
WhisperCool
®
WhisperCool
TM
chiller electrical
TM
chiller is