Chiller

Fill the reservoir up with the coolant fluid with inhibitor (Part No. N0776200). The
chiller reservoir has a capacity of 4.2 liters (1.1 US gallons).
Cooling water pressure: 310-550 kPa (3.1 to 5.5 bar or 45-80 psig)

Vent

Check that your vent system is switched on and is not blocked. Check regularly for
proper flow rate of the vent system (120 CFM/min).

Torch and Plasma Induction Plates

Inspect the torch, glassware, and aerosol injector tube. The glassware should be
clean, with no traces of deposits or signs of melting.
o extend the life and aid in preventing devitrification of glassware, it is
T
Caution
necessary to remove all traces of alkali on glassware caused by handling it
with bare hands. To remove the alkali, wipe the glassware clean with
cotton wool dipped in alcohol.
Pour prolonger la durée de vie et de l'aide dans la prévention de la
Attention
dévitrification de la verrerie, il est nécessaire d'éliminer toute trace d'alcali
sur la verrerie provoquée par la manipulation à mains nues. Pour supprimer
l'alcali, essuyez la verrerie propre avec un coton imbibé d'alcool.
Daily Checks