Removing And Disassembling The Torch - PerkinElmer AVIO 500 Customer Hardware And Service Manual

Icp optical emission
Table of Contents

Advertisement

3. Rotate the injector adapter to the left, then pull gently down - when the locking
pin is out and you apply pressure, the whole torch body could come out.
1
Figure 5-2
Item
1
2
3

Removing and Disassembling the Torch

1. Turn off the plasma if it is on.
If the plasma has been on, the torch glass will be very hot and can
!
cause serious burns. Wait five minutes after turning off the plasma
before you begin these maintenance procedures.
Warning
Avertissement: Si le plasma a été mis sur le verre de la torche sera
très chaude et peut causer des brûlures graves. Attendez cinq minutes
après avoir éteint le plasma avant de commencer ces procédures
d'entretien.
Quick-Change Adjustable Torch Module
Removing the spray chamber assembly and injector.
Scott spray chamber
Injector lock
Injector
Description
209
2
3

Advertisement

Table of Contents
loading

Table of Contents