Connecting The Polyscience Chiller - PerkinElmer AVIO 500 Customer Hardware And Service Manual

Icp optical emission
Table of Contents

Advertisement

An air hose with 1/4-in. Swagelok fittings at each end is supplied in the hose kit
that is shipped with the instrument. A second air hose is included with the
PerkinElmer Air Dryer Filter in the Installation kit (Part No. N0770432). The
following procedure describes how to connect the PerkinElmer air compressor,
filter and regulator.
If you are using house air, it must be clean and dry. The instrument
Note
includes an air dryer assembly (Part No. N0775325). Refer to Connecting
the Shear Gas Supply later in this chapter.
1. Install the Air Dryer Filter as described in the instructions that are included
with the filter.
2. Locate the air hose (Part No. N0770348, uses 1/4-in. Swagelok fittings at each
end) that is shipped with the Air Dryer Filter. Connect the air hose from the air
compressor to the AIR IN fitting on the Air Dryer Filter using a 1/4-in. space
collar at both ends. Make sure the air flow is in the proper direction, as
indicated by arrows on top of the filters.
3. Locate the air hose (Part No. N0770348, uses 1/4-in. Swagelok fittings at each
end) that is shipped with the instrument in the hose kit. Connect the air hose
from the AIR OUT fitting on the Air Dryer Filter using a 1/4-in. space collar
at both ends.

Connecting the PolyScience Chiller

The PolyScience
the instrument. This chiller requires a single wall receptacle. The PolyScience
WhisperCool
TM
familiar with this manual before proceeding.
You are required to have your electrician install the single wall
Note
receptacle for the PolyScience
Connecting the Gases and Cooling Water
®
WhisperCool
TM
Chiller must be used to supply cooling water to
Chiller comes with a complete instruction manual. You should be
®
WhisperCool
TM
chiller.
121
®