GemCone Nebulizer

Daily Cleaning of the Low-Flow GemCone Nebulizer

Each day after use, the low-flow GemCone nebulizer should be rinsed out by
pumping deionized water to the nebulizer for several minutes. This should be done
with both the plasma on and the nebulizer gas on. This helps prevent the formation
of salt crystals near the exit of the nebulizer gas orifice. The plasma should be on
to prevent aerosol from condensing on the purge window.

Daily Cleaning of the High Solids GemCone Nebulizer

It is important to rinse the high solids nebulizer after use, particularly when
working with solutions containing high levels of solids. You do this by pumping
deionized water for several minutes to wash out any solid material that may have
crystallized on the inside surface of the nebulizer. If a stronger cleaning solution is
required, a solution of 1% acid in deionized water may be used in place of the
deionized water. The plasma should be on to prevent aerosol from condensing on
the purge window.
To rinse the argon orifice:
1. Place the sample capillary tube in deionized water or 1% acid in deionized
water and operate the pump.
2. Momentarily shut off the nebulizer gas flow for about 30 seconds.
3. Turn the nebulizer gas back on.
This procedure can be performed more than once a day in cases where very
high levels of dissolved solids are aspirated through the nebulizer for
prolonged periods of time.