Chapter Overview

The Avio 500 is a complete system, with the exception of the following items:
suitable working area, exhaust vent, gases and regulators, cooling water, and a
computer table or bench. The items shown in the following checklist need to be
considered when preparing the laboratory for the instrument:

Environmental Conditions

Exhaust Vent Requirements
Laboratory Space Requirements
Cooling Water Requirements
Electrical Requirements
Pneumatic Requirements
Environmental Conditions
The laboratory in which the Avio 500 Spectrometer system is located must meet
the following conditions:
Locate the instrument in an area free of corrosive fumes and excessive
vibration.
The instrument will operate with a laboratory temperature between 15 and 35
°C (59-95 °F) with a maximum rate of change of 2.8 °C (5 °F) per hour. For
optimum instrument performance, the room temperature should be controlled
at 20
Preparing Your
Laboratory
± 2 °C.