Connections To Electrical, Gas, And Cooling Water Supplies - PerkinElmer AVIO 500 Customer Hardware And Service Manual

Icp optical emission
Table of Contents

Advertisement

System Description
106

Connections to Electrical, Gas, and Cooling Water Supplies

All gas lines supplied with the instrument are 6.5m (20 ft.) long. Ensure
Note
that the gas connections are located in close range (within 6.5 m) to the
instrument.
Electrical
Connections
Gas Connections
Label reads...
(right side of
instrument as viewed
from the front of the
instrument)
Ethernet connection
(on left side of
instrument as viewed
from the front of the
instrument)
N
Supply
2
Shear Gas Supply
Ar Supply
Use...
To connect...
Power cable
Instrument to 200-
(attached to
240 VAC, 16 amp
instrument)
IEC-309 connector
outlet
Ethernet cable
Instrument to
(Part No.
computer
09290332)
Nitrogen hose
Nitrogen inlet to
(Part No.
nitrogen supply
N0690275)
Air hose
Require a supply of
(Part No.
air (typically) or
N0770348)
nitrogen for the
included with
shear gas.
Air Dryer Filter
and regulator
assembly
Argon hose
Argon inlet to argon
(Part No.
supply
N0690274)