Figure 3-8.
The MEINHARD nebulizer will aspirate liquid sample naturally. The
Note
nebulizer may also be operated with externally pumped sample, provided that
the sample flow is not significantly below the natural aspiration rate.

Peristaltic Pump

The peristaltic pump is fully computer-controlled. As an added feature, the pump
speed can be programmed to run at a fast speed for the read or rinse cycle. The
TubingSaver mode is a feature for extending the life span of pump tubing.

Autosampler

PerkinElmer offers autosamplers for automated sample handling. These
autosamplers consist of a sample table, a sample tray, and a motorized sampling
arm with an attached probe. Different sample trays are available for each
Concentric glass nebulizer and end cap N0680504 (also shown in
cross-sectional view).
Sample Introduction System