Installation
126
Remote Chiller Operations
The remote chiller will automatically be on when the Avio 500 Spectrometer is
warming up. The warming up period includes just being turned on, coming out of
Sleep mode or Standby mode.
If the instrument has just been turned on, starting the software will automatically
begin warm up. Initiating the software will have no effect on the chiller if the
chiller is already on.
If the Avio 500 Spectrometer is in Sleep or Standby mode and the instrument is still
ignited, the chiller will remain on. If the instrument is not ignited in Sleep or
Standby mode the chiller will turn itself off.
The plasma should never be ignited if the chiller is off.
Note

Filling the PolyScience Chiller

Use the following procedure to fill the chiller.
1. Turn the filler cap counter clockwise and lift the cap off.
2. Fill the reservoir with the coolant fluid with inhibitor (Part No. N0776200).
The chiller reservoir has a capacity of 4.2 liters (1.1 US gallons).
3. Keep some coolant fluid beside the unit and proceed to Starting Up the
PolyScience Chiller. The reservoir tank may need to be topped off once the
unit has been started.