Scott Spray Chamber; Removing The Scott Spray Chamber - PerkinElmer AVIO 500 Customer Hardware And Service Manual

Icp optical emission
Table of Contents

Advertisement

Maintenance
222
When installing the Torch Module in the next steps, do so carefully so you will
avoid breaking the torch and risking possible injury.
!
Avertissement: Lors de l'installation du module de la flamme dans les
Warning
prochaines étapes, faire avec soin afin que vous éviter de casser la torche et un
risque de blessure possible.
Wear cotton gloves when you handle the torch glassware.
Caution
Attention
Porter des gants de coton lorsque vous manipulez la verrerie de la torche.
7. To install the torch see Installing the Quick-Change Adjustable Torch Module
on page 132.

Scott Spray Chamber

Check the spray chamber periodically for leaks around the nebulizer end cap and
drain fitting. Examine the end cap O-ring for cracking and wear and replace if
necessary. Deposits and plated out metals can accumulate in the spray chamber,
which can introduce "memory effects." Usually, aspirating a strong acid or solvent
will clean out these deposits. If not, remove the spray chamber and clean it in a 5%
to 20% nitric acid bath

Removing the Scott Spray Chamber

1. Place the sample tubing in deionized water and turn on the pump to flush the
spray chamber. Remove the sample tubing from the deionized water and
continue to run the pump a few minutes until the spray chamber is fully
drained. Then turn off the pump.
2. Disconnect the spray chamber drain tubing from the pump tubing for the drain.
Be careful not to have liquid spill out of the disconnected drain tubing. Spilled
liquid should be cleaned up immediately. Check that the tubing is in good
condition and replace if necessary

Advertisement

Table of Contents
loading

Table of Contents