Troubleshooting
328
Table 6-16.
Problem
Possible Cause(s)
Cannot get
The standard may have too
suggested
high a concentration.
detection
limits
(continued)
You may not be comparing
the appropriate value for the
wavelength you are using.
General
Performance problems such
performance
as poor precision or loss of
problems
signal can often be traced to
the setup and condition of the
sample introduction system.
You are having difficulty
getting values for silicon in
samples with HF.
Unexplained
The torch or injector may be
background
coated with a deposit that
shift
reflects light, resulting in a
periodic background shift.
Plasma Ignition and Stability Problems
Plasma Ignition Problems
Possible reasons for unsuccessful ignition follow. Correct the problem, if
possible, and then try to ignite the plasma again.
•
If the plasma will not light check the ignitor. Open the door to the torch
compartment. Check that the ignitor contact finger is making contact with the
Performance Problems: Troubleshooting Table
Corrective Action
The standard should be 100 times the
estimated detection limit.
Confirm that you are using the appropriate
wavelength and recheck the value. Refer to
the Syngistix Software Guide on the
Syngistix Document Pack CD for details..
See the section Checking the Sample
Introduction System on page 320.
Use a sapphire injector.
Replace the torch and injector.