Troubleshooting Tasks Performed; Common Setup Problems - PerkinElmer AVIO 500 Customer Hardware And Service Manual

Icp optical emission
Table of Contents

Advertisement

Troubleshooting Tasks Performed:

Did the plasma light:
Did the plasma stay lit after ignition:
Plasma went out after pump turned on:
Plasma went out after how long of operation:
Plasma went out with Dry Plasma (neb on, no pump):
Plasma went out when injector capped off:
Referring to the figures on the following pages, the plasma looked like

Common Setup Problems:

Glassware bottomed out in torch body (Glassware Holder):
Glassware aligned with notch in torch body (Glassware Holder):
Igniter finger contacting glassware:
Torch position appropriate for sample type:
Aqueous = -3, Organics = -3, High Solids = -1
Injector O-Rings not worn, cracked, or cut:
Torch Body (Glassware Holder) O-Rings not worn, cracked, or cut:
Injector Dry:
Spray Chamber draining properly:
Pump Tubing clamped:
Pump Tubing not worn out:
Pump Tubing pumping in proper direction:
Radial and Axial windows installed and clean:
RF Generator/Sample Introduction System Troubleshooting Form
335
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
:
FIGURE
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No