Chiller and Filters

Inspect the metal strainer (Part No. N0776075) that is suspended inside the chiller
reservoir; this should be removed and rinsed clean when the system is flushed or if
dirty. Flush out the chiller every 6 months.
Fill the chiller reservoir with the coolant fluid with corrosion inhibitor (Part No.
N0776200). The chiller reservoir has a capacity of 4.2 liters (1.1 US gallons). Refer
to Connecting the Gases and Cooling Water on page 118 for more information on
filling the chiller.
3
3
Location of the air filters on the left and right side of the instrument.
Left Side of the Instrument
Right Side of the Instrument
Air Filters (Part No. 09995098)