Nebulizers - PerkinElmer AVIO 500 Customer Hardware And Service Manual

Icp optical emission
Table of Contents

Advertisement

Maintenance
254
8. Replace the cover and return the assembly to the instrument.
9. Reattach the Axial Flat Optic Temperature Sensor to the instrument.
10. Return the thumbscrew on the heat sink cover.
11. Return the Axial Heat Sink Cooling Tube. See Remove and Replacing the
Axial Heat Sink Cooling Tube on page 247.

Nebulizers

PerkinElmer offers different types of nebulizers for a wide variety of applications.
For descriptions of the nebulizers and their different uses and advantages, refer to
Chapter 3, System Description. Part numbers and replacement parts for the
different types of nebulizers are listed on the next page.
Recommended flow rates for the nebulizers are given in the table below. If you find
that a flow rate is insufficient, this often indicates that the nebulizer has clogged
and requires cleaning. On GemTip and GemCone nebulizers, you should also
check that the argon gas fitting is properly tightened. If a space collar is used,
tighten the fitting with a wrench until you can no longer turn the knurled space
collar. If a space collar is not used, finger-tighten the fitting, then using a wrench,
turn clockwise 1/8-turn.
Table 5-5.
Nebulizer
GemTip Cross-Flow
Low-Flow GemCone
High Flow GemCone
Nebulizer Flow Rates
Typical Value
0.6 L/min
0.6 L/min
1.0 L/min
Argon Flow Rate
Operating Range
0.4 - 1.3 L/min
0.4 - 0.95 L/min
0.7 - 1.0 L/min