SoundAdvisor Model 831C
Selecting Single Block Timer Mode With Measurement History
a.
Enter a Start (y-m-d) date and Time, and an End (y-m-d) date and
Time, where Time is the time of day.
b.
Select the Enable Measurement History checkbox.
c.
If you want to send a measurement interval exceedance alert, select
your preferred method in-line with Alerts, and select a Trigger Source
and Trigger Level (dB) for the alert. The Trigger Level cannot be less
than 0.1 (dB).
Trigger Sources for this alert include the following statistics:
TAKE NOTE
LAeq, LCeq, LZeq, Lmax, Lmin, and all 6 values of Ln. If the current dB level is
greater than the Trigger Level (dB) of the selected Trigger Source at any time
during the current measurement interval, the 831C sends an alert notification
via email or text when the interval ends.
d.
Select any additional settings on the Setup Manager tabs. For more
information, see 6.2.2 Customizing Setup Values On the Setup
Manager Tabs.
e.
Press Close (
f.
Set it as the Active setup file.
Selecting Daily Timer Mode With Measurement History
a.
Enter a Start (y-m-d) date and Time, and an End (y-m-d) date and
Time, where Time is the time of day.
b.
If desired, select the checkbox in-line with 2, and 3 to enter additional
measurement periods.
c.
Select the Enable Measurement History checkbox.
d.
If you want to send a measurement interval exceedance alert, select
your preferred method in-line with Alerts, and select a Trigger Source
and Trigger Level (dB) for the alert. The Trigger Level cannot be less
than 0.1 (dB).
Trigger Sources for this alert include the following statistics:
TAKE NOTE
LAeq, LCeq, LZeq, Lmax, Lmin, and all 6 values of Ln. If the current dB level is
greater than the Trigger Level (dB) of the selected Trigger Source at any time
during the current measurement interval, the 831C sends an alert notification
via email or text when the interval ends.
e.
Select any additional settings on the Setup Manager tabs. For more
information, see 6.2.2 Customizing Setup Values On the Setup
Manager Tabs.
f.
Press Close (
g.
Set it as the Active setup file.
Setting Up a Measurement With Measurement History
/*/
) and save your Setup.
/*/
) and save your Setup.