Reference Graphs and Tables

Increase in Measured Noise as Ref Source Approaches UUT Noise

Increase in Measured Noise as Ref Source
Approaches UUT Noise
The graph shown above demonstrates that as the noise level of the reference
source approaches the noise level of the UUT, the level measured by the
software (which is the sum of all sources affecting the test system) is
increased above the actual noise level of the UUT.
Agilent Technologies E5500 Phase Noise Measurement System 18-5