Power Requirements

Table 2-1
Instrument
85101C display processor
85102B IF detector
83621B synthesized source (LO)
83651B synthesized source (RF)
E7341A millimeter-wave controller
E7342L left test head (85 GHz)
or E7352L left test head (110 GHz)
E7342R right test head (85 GHz)
or E7352R right test head (110 GHz)
Total
Installation

Site Preparation

Site Preparation

Before installing the system, be sure that the required ac power is available
at all necessary locations.
Three-wire power cables (which provide a safety ground) must be used
with all instruments.
Air-conditioning equipment (or other motor-operated equipment) should
not be placed on the same ac line that powers the system.
The table below lists the maximum VA ratings and BTU/hour ratings for
all instruments in the system. This table can be used to determine both
the electrical requirements and the air conditioning requirements of the
system.
Power requirements of the system
Standard Equipment
Maximum VA Rating
250
210
400
400
500
(powered from controller)
(powered from controller)
1760
NOTES:
(1) Values are based on 120 Vac supplied to each instrument at 60 Hz.
(2) The millimeter-wave controller supplies power to the test heads.
2-2 8510XF Network Analyzer Systems
Maximum BTU/hour
850
714
1360
1360
1700
(powered from controller)
(powered from controller)
5984