Other Connections
and Settings
GPIB Addresses
Table 2-13
Language Switches
When the 8510C power is turned on, all previously assigned GPIB addresses
are automatically recalled from memory and assigned to the
various system instruments, including the address of the 8510C itself.
The GPIB address switch settings for all instruments must match the
addresses assigned by the system. In 8510XF systems, these addresses are
set at the factory prior to shipment. For reference, the instruments and the
applicable default addresses are listed in the table below.
8510XF System GPIB Addresses
Instrument
8510C
System Bus
83621B LO Source
83651B RF Source
Millimeter-wave controller
Plotter
Printer
Disk Drive
The LANG switches of the 83621B and 83651B sources must be set to 111.
GPIB Address
16
17
18
19
23
5
1
0
8510XF Network Analyzer Systems 2-37
Installation
Cable List