Agilent Technologies 8510XF Operating And Service Manual page 97

Network analyzer systems 2 to 85 ghz and 2 to 110 ghz
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

Full 2-Port Isolation
Standards
Figure 3-15
Notes on the Standards
NOTE
These standards are used to measure forward and reverse isolation for full
2-port calibrations. When you perform a calibration of that type using a
1.0 mm calibration kit, you will see the following menu:
Full 2-Port Isolation Calibration Menu & Equivalent GPIB Commands
{FWD ISOL'N
Refer to "Isolation" under "Banded Standards" in
LOAD}
Table 3-1 on page
{REV ISOL'N
Refer to "Isolation" under "Banded Standards" in
LOAD}
Table 3-1 on page
To skip the isolation part of the calibration, select
3-12.
3-12.
{OMIT ISOLATION}
8510XF Network Analyzer Systems 3-27
Operation
Standard Types
.