Agilent Technologies 8510XF Operating And Service Manual page 93

Network analyzer systems 2 to 85 ghz and 2 to 110 ghz
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

Load/Short Standards
Figure 3-11
Notes on the Standards
These standards are used to measure reflection for:
S11 1-Port calibrations
S22 1-Port calibrations
One-Path 2-Port calibrations
Full 2-Port calibrations
When you perform any of these calibrations using a 1.0 mm calibration kit,
you will see the following menu:
Load/Short Standard Menu & Equivalent GPIB Commands
Refers to either of two 50 GHz loads (with male and
{50GHz LOAD}
female connectors, as appropriate). This standard is used
for frequencies up to 50 GHz.
Refers to Short No. 4 with a male connector (that is, the
{(F) SHORT4}
short is to be connected to a female port). This standard is
used for frequencies from 50 to 75 GHz.
Refers to Short No. 2 with a male connector (that is, the
{(F) SHORT2}
short is to be connected to a female port). This standard is
used for frequencies above 75 GHz.
Refers to Short No. 4 with a female connector (that is, the
{(M) SHORT4}
open is to be connected to a male port). This standard is
used for frequencies from 50 to 75 GHz.
Refers to Short No. 2 with a female connector (that is, the
{(M) SHORT2}
short is to be connected to a male port). This standard is
used for frequencies above 75 GHz.
Standard Types