Full 2-Port
Figure 3-7
Notes on the Standards
Operation
Calibration Types
This is a very thorough calibration, which takes measurements of reflection,
transmission, and isolation for both ports. This type of calibration (when
performed using the 1.0 mm calibration kit) combines the SOLT calibration
technique (for frequencies up to 50 GHz) with the Offset Shorts calibration
technique (for frequencies above 50 GHz). This combination yields the best
possible calibration results, for measurement over the full frequency range
of the 8510XF.
For this calibration, select:
or
[CAL] {CAL1
CAL2} {FULL 2-PORT}
This calls up the following menu:
Full 2-Port Menu & Equivalent GPIB Commands
These are the same standards that are used for the S11
{REFLECT'N}
and S22 1-port calibrations, except that in this case both
ports are calibrated. The reflection standards are sorted
into three categories:
See
"Open/Short Standards" on page 3-21
See
"Shorts Standards" on page 3-22
See
"Load/Short Standards" on page 3-23
{TRANS-
See
"Full 2-Port Transmission Standards" on page
MISSION}
See
"Full 2-Port Isolation Standards" on page
{ISOLATION}
3-18 8510XF Network Analyzer Systems
3-26.