Agilent Technologies 8510XF Operating And Service Manual page 141

Network analyzer systems 2 to 85 ghz and 2 to 110 ghz
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

NOTE
If a connector appears defective, clean and inspect it again. Damaged
connectors should be discarded or sent for repair. Try to determine the cause
of damage before connecting a new, undamaged connector in the same
configuration.
Cleaning connectors
1. Inspect the connectors for dirt or debris.
2. Blow off any accumulated dust, using clean (filtered) compressed air.
3. Dip the cleaning swab in isopropyl alcohol.
Do not pour alcohol directly on the parts to be cleaned, and do not allow
alcohol to wash down inside the connectors. An excessive flow of alcohol
may carry residue into the interior of the connectors.
4. Gently wipe all connecting surfaces with the end of the cleaning swab.
5. Dry with compressed air.
6. Inspect the connectors again, and repeat the cleaning procedure if
necessary.
Connector care supplies
The following supplies can be ordered from Agilent Technologies:
Item Description
Magnifying lens (10x)
Isopropyl Alcohol (30 ml, 99.5%)
Cleaning Swabs (100 each)
Torque Wrench (6 mm, 4 in-lb)
Open End Wrench (6 mm)
System Maintenance
1.0 mm Connector Care
Part Number
1000-1114
(included in Calibration Kit)
8500-5344
9301-1243
8710-2079 (included in Calibration Kit)
8710-2156 (included in Calibration Kit)
8510XF Network Analyzer Systems 5-5