Preparation For Power Supply; Power Requirements; Power Cable - Agilent Technologies ENA Series Service Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Power Requirement

Preparation for Power Supply

Preparation for Power Supply
Before turning on power to the equipment, be sure to verify the following:

Power Requirements

The E5061A/E5062A requires the following power source:
Voltage
Frequency
Maximum power consumption

Power Cable

In accordance with international safety standards, this instrument is equipped with a
three-wire power cable. When connected to an appropriate ac power outlet, this cable
grounds the instrument frame. The type of power cable shipped with each instrument
depends on country of destination. Refer to Figure E-1 for the part numbers of the power
cables available.
WARNING
For protection from electrical shock, the power cable ground must not be defeated.
The power plug must be plugged into an outlet that provides a protective earth
ground connection.
234
*1. Switched automatically by the E5061A/E5062A in conformity to the voltage.
Requirements
90 to 132 VAC or 198 to 264 VAC
47 to 63 Hz
350 VA
*1
Appendix E