Broadband Standards

Alternative 1.0 mm Calibrations

The two calibration techniques described here are simplified broadband
calibrations, which save time at the expense of accuracy. These calibrations
are meant to be used only in situations where measurement accuracy is not
critical.
The broadband calibrations use a separate set of calibration standard
definitions, which are included on the calibration data disk that is supplied
with the 1.0 mm calibration kit (85059A).
Loading the broadband standards
1. Insert the calibration data disk into the 8510C disk drive.
2. Press
.
[DISC] {LOAD}
The analyzer displays
3. Select
or
{CAL KIT 1}
{CAL KIT 2}
The analyzer displays
4. Select
from the display menu. (This is the file that contains
CK_1MMA1B
the broadband standards.)
5. Select
. This loads the calibration constants into the memory
{LOAD FILE}
of the network analyzer. A screen message indicates when the load
process is complete (
6. Remove the disk from the drive.

Alternative 1.0 mm Calibrations

SELECT DATA TYPE TO LOAD
.
USE KNOB OR STEP KEYS TO SELECT A FILE
FILE 'CK_1MMA1B' LOADED FROM DISK
8510XF Network Analyzer Systems 3-43