Agilent Technologies 8510XF Operating And Service Manual page 125

Network analyzer systems 2 to 85 ghz and 2 to 110 ghz
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

3. A screen prompt will give you the opportunity to set the date and time.
Press the
and time.
4. The program loads the System Configuration file, and displays the
System Hardware Configuration Menu. Use this menu to specify the
equipment you are using, and the frequency range you are verifying
(refer to
To setup for the first performance verification, select the proper
hardware setup for Low Band Verification (< 50 GHz) using
on page 4-4
for the E7342A 85 GHz system.
a. Select
menu item.
b. Use the up and down arrow keys on the PC to advance from one
menu item to another.
c. Select
5. The main menu is displayed (see
Table 4-4
Main Menu
Softkey
System Config
System Specs
System Uncert
Edit Specs
Verify System
Quit Progam
key on the PC to skip this step, or press
[Y]
"Frequency ranges" on page
for the E7352A 110 GHz system or
or
to step through the list of selections for each
NEXT
PREVIOUS
when you have made a selection for each menu item.
DONE
Description
System Hardware and Software Configuration Menu
System Spec Table Menu
System Uncertainty & Dynamic Accuracy Plot/Table Menu
Edit the System Specifications
System Performance Verification Menu
Quit this Program
8510XF Network Analyzer Systems 4-7
Performance Verification
Verification Setup
to set the date
[N]
4-4).
Table 4-3 on page 4-4
Table
4-4).
Table 4-2