Agilent Technologies 8510XF Operating And Service Manual page 131

Network analyzer systems 2 to 85 ghz and 2 to 110 ghz
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

12. When verification has been completed view the data, then print or save
(if required). If needed the verification standard can be remeasured by
selecting:
Repeat Measure
13. Select
to return to the Standard Selection Menu. Repeat
Select Standard
steps 7
through
12
above for the other verification standard (refer to
Table 4-6 on page
4-10).
14. Select
until the Main Menu appears.
Prior Menu
15. Select
to exit the program, then close the program window to
Quit Program
finish.
Performance Verification
Verification Procedures
.
8510XF Network Analyzer Systems 4-13