Installation

Turning On the System

Turning On the System

While turning on the system, be sure to observe all of the precautions stated
on
page
2-9.
1. Verify that the hardware is properly connected (refer to
Configurations" on page
2. Inspect the test ports before they are connected to any other device (see
"Visual inspection" on page 5-4
in a 1.0 mm connector.)
3. Turn on power for the system rack.
4. Turn on power for all system instruments, except the network analyzer
display unit (the upper portion of the 8510C).
5. Wait 10 seconds, then turn on the network analyzer display unit.
6. The 8510C begins making measurements according to the last recording
of instrument state #8. (Instrument State 8 is recalled each time line
power to the network analyzer is turned on, and each time
on the 8510C is pressed.)
2-38 8510XF Network Analyzer Systems
2-17).
for information on recognizing defects
"Basic System
[USER PRESET]