Table 7-5
MESSAGE
RPG INACTIVE FOR STEP MODE STIMULUS
STEP KEYS INACTIVE FOR STEP MODE STIMULUS
FUNCTION NOT VALID WITHOUT SYSTEM PWR LEVELING
(GPIB Status Message: 206)
CANNOT POWER SWEEP WITHOUT SYSTEM PWR LEVELING
(GPIB Status Message: 202)
FUNCTION NOT VALID IN SYSTEM PWR LEVELING
(GPIB Status Message: 203)
GAIN CAL (or CONVERSION LOSS CAL)
LOAD/SAVE FROM TESTSET EEPROM FAILED
(GPIB Status Message: 204)
LVL DET GAIN CAL PROBLEM (RANGE/RESOLUTION)
(GPIB Status Message: 207)
SUBSET CANNOT CONTAIN MORE THAN 801 PTS
REQUESTED SWEEP MODE NOT SUPPORTED
PORT 1, 2 to 50 GHz MIXER NOT DETECTED
PORT 2, 2 to 50 GHz MIXER NOT DETECTED
PORT 1, V BAND MIXER NOT DETECTED
PORT 2, V BAND MIXER NOT DETECTED
Menus & Commands
New Messages
Warnings messages
The following warning messages (which are displayed in red letters on the
screen of the 8510C) were created specifically for the 8510XF firmware.
Warning Messages (Text Displayed in Red)
7-50 8510XF Network Analyzer Systems
NOTES
The "RPG" knob cannot be used in connection with this
function.
The step keys cannot be used in connection with this
function.
These functions can be used only if SYSTEM leveling is
selected on the [SYSTEM] {MORE} {RF POWER CONFIG}
menu (see
"RF Power Configuration" on page
This function cannot be used if SYSTEM leveling is selected
on the [SYSTEM] {MORE} {RF POWER CONFIG} menu
(see
"RF Power Configuration" on page
The GAIN CAL (or CONV LOSS CAL) data cannot be
loaded from the EEPROM in the millimeter wave controller
(however, no functionality is disabled).
Detector gain cal routine failed; indicates hardware failure or
improper installation of system or cabling. No functionality is
disabled.
Equivalent message in standard 8510C firmware reads "792
PTS".
8510XF does not support ramp mode.
These hardware failures are detected by a self-test routine,
which is run at power-up or when the test set address is
changed.
3-30).