LO Power Settings

LO Power On/Off
LO Power Settings
NOTE
LO Source Power
LO Source Slope
If
is selected, the LO source performs its own leveling,
{LO LEVEL / EXTERNAL}
using an external detector.
If
{LO LEVEL / LEVELING OFF}
mode.
The
softkey calls up a submenu (see
{LO POWER SETTINGS}
Settings"
below).
The softkeys
{LO POWER /ON}
source power.
[SYSTEM] {MORE} {LO POWER CONFIG} {LO POWER SETTINGS}
which is shown on the right of
The functions represented on this submenu will not work if
is selected on the
SYSTEM}
Power Configuration" on page
normal operation of the system; they are made available for service use, or
for unusual measurement situations.
, followed by a number, specifies a power level setting for
{LO SOURCE POWER}
the 83621B LO source, in dBm.
and
{SLOPE ON}
{SLOPE OFF}
LO source.
, followed by a number, specifies the power slope
{SLOPE ON}
setting for the LO source, in dB/GHz.
is selected, the LO source is set to the unleveled
and
activate and deactivate LO
{LO POWER /OFF}
Figure 3-20
on the previous page.
[SYSTEM] {MORE} {LO POWER CONFIG}
3-36). These functions are not used during
are used to enable and disable power slope for the