Agilent Technologies 8510XF Operating And Service Manual page 89

Network analyzer systems 2 to 85 ghz and 2 to 110 ghz
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

TRL (TRM) 2-Port
NOTE
Figure 3-8
Notes on the Standards
The 1.0 mm calibration kit does not include a precision transmission line;
therefore a TRM (Thru-Reflect-Match) calibration must be used in place of
the TRL (Thru-Reflect-Line) calibration indicated by the menu title.
The 1.0 mm calibration kit only supports TRM 2-port calibration for
measurement up to 50 GHz.
For this calibration, select:
or
[CAL] {CAL1
CAL2} {TRL 2-PORT}
TRL 2-Port Menu & Equivalent GPIB Commands
Connect Port 1 to Port 2.
{THRU}
{S11 REFLECT
Connect a short to Port 1. You will be prompted to select
SHORT}
{(F) SHORT3}
{S22 REFLECT
Connect a short to Port 2. You will be prompted to select
SHORT}
{(F) SHORT3}
{ISOLATION
See
"Full 2-Port Isolation Standards" on page
LOAD}
{LINE MATCH
Connect 50 GHz loads to Port 1 and Port 2.
MATCH}
{LOWBAND
This standard is not used in connection with the 1.0 mm
REFLECT'N}
calibration kit (for information regarding its use with
other calibration kits, see the manuals for those kits).
or
.
{(M) SHORT3}
or
.
{(M) SHORT3}
8510XF Network Analyzer Systems 3-19
Operation
Calibration Types