Agilent Technologies 8510XF Operating And Service Manual page 85

Network analyzer systems 2 to 85 ghz and 2 to 110 ghz
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

Response & Isolation
Figure 3-4
Select ONE of the
Response Standards
Isolation Standard
This is a very limited calibration, which corrects only for frequency response
and isolation, not for directivity or source match errors. The performance of
the 8510XF is not specified, if this calibration is used. To perform this
calibration using the 1.0 mm calibration kit, select:
or
[CAL] {CAL1
CAL2} {RESPONSE & ISOL'N}
This will call up the menu and submenu shown below:
Response & Isolation Menu, Response Menu, and Equivalent GPIB
Commands
For Reflection Measurements (S11, S22), select one of the following:
Refers to offset short No. 3 with a male connector (that
{(F) SHORT3}
is, the short is to be connected to a female port).
Refers to an open with a male connector (that is, the open
{(F) OPEN}
is to be connected to a female port).
Refers to offset short No. 3 with a female connector (that
{(M) SHORT3}
is, the short is to be connected to a male port).
Refers to an open with a female connector (that is, the
{(M) OPEN}
open is to be connected to a male port).
For Transmission Measurements (S12, S21), select:
Connect Port 1 to Port 2 by way of a cable.
{THRU}
{ISOL'N STD
Refer to "Isolation" under "Banded Standards" in
LOAD}
Table 3-1 on page
Calibration Types
3-12.
8510XF Network Analyzer Systems 3-15