Disinfecting; Sterilizing; Removing/Fitting Probe Holder (Devices With Temperature Probe); Removing The Probe Holder - Seca mVSA 535 Instructions For Use Manual

Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

8.2 Disinfecting

Component (depending on equipment)
: Monitor with
seca mVSA 535
SmartBucket
Blood pressure cuff and compressed air
tube
Temperature probes (red/blue) with cable
Probe holder (red/blue)
SpO
sensor with cable
2
Patient cable for SpO

8.3 Sterilizing

8.4 Removing/fitting probe holder (devices with temperature probe)

Removing the probe holder

82 •
seca mVSA 535
1. Follow the instructions for use for the disinfectant.
2. Disinfect the device and its accessories as described in the table:
Interval
As required
As required
As required
As required
Follow the manufacturer's instructions for use
sensor
Follow the manufacturer's instructions for use
2
The device and accessories must not be sterilized.
1. Open the covering cap.
1. Remove all the measuring accessories (measuring
devices and consumables) from the device (depending
on version)
• ➔
Removing the probe holder
2. Moisten a soft cloth with disinfectant (active ingredient:
quaternary ammonium compounds)
3. Wipe over all surfaces
4. Allow to air-dry for approx. 30 minutes
1. Moisten a soft cloth with disinfectant (active ingredient:
70 % isopropanol)
2. Wipe over cuff and compressed air tube
3. Rinse thoroughly with water
4. Allow to dry at room temperature
1. Discard probe cover and dispose of it
2. Moisten a soft cloth with disinfectant (active ingredient:
70 % isopropanol)
3. Wipe over temperature probe
4. Shake out temperature probe so that no liquid remains
in it
5. Allow to air-dry for approx. 30 minutes
1. ➔
Removing the probe holder
2. Moisten a cotton bud with disinfectant (active
ingredient: 70 % isopropanol)
3. Wipe over the inner surfaces of the probe holder
4. Allow to air-dry for approx. 30 minutes
WARNING!
Patient hazard
The color of the probe holder indicates whether a device is intended
for oral/axillary or rectal temperature measurement. This distinction
can no longer be made once the probe holder has been removed.
Confusing the probe holders can lead to cross-contamination.
► Ensure that the probe holder is fitted back in the device from which
it was removed following a hygiene treatment.
Disinfecting