Functions
PC
ComStation
Inspection interval: 3 months
Are the PC, screen, UPS, speech unit and
mouse undamaged and free of contami-
nation?
Are the connection cables undamaged
and free of contamination?
Are all plugs of the connection cables
seated firmly and are the screws screwed
in?
Is the screen easily readable?
Is a fault in the nurse call system shown
on the screen? If so, locate the fault and
remedy it.
Are calls of the ward indicated by a call
tone and shown on the screen?
Inspection interval: 12 months
Can the call be answered?
For calls with speech possibility: Is a
speech connection to the call location es-
tablished when a call is answered?
Can both conversation partners hear at an
appropriate volume and understand each
other? Test using the microphone and
loudspeaker and test using the handset.
For the "Call" call type: Can you cancel the
call remotely?
For the "Door Call" call type: Can you trig-
ger the door opening mechanism to let the
person at the door in?
For other call types: Can you close the
speech connection to the call location?
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)
Order no.
77 0602 00
14. Servicing