14. Servicing
14.6

Connection sockets

Functions
Connection socket with call switch
Inspection interval: 3 months
 Is the housing undamaged and free of
contamination?
 Can a call be triggered by pressing the red
call button?
 Do the location and reassurance lights of
the red call button work?
 Do the plug sockets sit precisely and firm-
ly?
 Is the plug connection stable when a plug-
gable call device (e.g., pear push switch)
is connected to the plug socket? Check
both plug sockets.
 Does the pluggable call device (e.g., pear
push switch) work when it is connected to
the connection socket? Test as described
for the pluggable call device.
 If present: Does the external call device
connected to the rear input of the connec-
tion socket work? Test according to the
documentation for the external call de-
vice.
Connection socket with call switch,
bedhead unit
Like 70 0171 03 A.