4. Interfaces
4.1

Interfaces in the room

4.1.1
Lighting control
The patient handset (74 0747 00) is equipped with 2 light switches to switch the read-
ing light and room light. The pear push switch with call and light switch provides one
switch to switch the light. For information on selecting the light relays refer to
chapter 12. "Light control" on page 65.
4.1.2
Entertainment/television
The patient handset (74 0747 00) can transmit entertainment and TV sound through
the integrated loudspeaker or connected headphones. The open sound is switched
off when the headphone cord is inserted.
Up to 5 installed entertainment programmes are connected via connection socket
combi or connection socket combi, bedhead unit. Switching on/off, programme se-
lection and setting the volume can be performed using the buttons of the patient
handset.
The IR TV control module universal (77 0360 11) enables the control of a television
(e.g., from Samsung, Philips or LG). The TV sound is transmitted to the patient hand-
set. Switching off, programme selection and setting the volume can be performed us-
ing the buttons of the patient handset.
The TV audio amplifier (77 0365 00) enables the transmission of the sound of any
television (with headphone jack) via the patient handset. In the case of this applica-
tion, the TV is not controlled with the patient handset.
Tunstall itself offers televisions (Hospital LED TV, professional) that can be con-
trolled by the patient handset in connection with the RAN interface (77 0840 00). The
TV sound is transmitted through the patient handset. Switching on/off, programme
selection and setting the volume can be performed using the buttons of the patient
handset.
4.1.3
Call devices
In addition to the wired call devices, wireless call devices can also be used. Radio
Receiver T (Z 00 8202 33) can be connected to the connection sockets in the socket
of the pear push switch. Radio Receiver-T receives the signals of the radio trigger
MyAmie (P68007/02), fall detector iVi (P68005/47) and other radio transmitters from
the Tunstall Telecare portfolio.
Note: The radio transmission is not monitored. The radio transmitters must therefore
be used only as additional call devices in connection with the nurse call system.
The RAN interface (77 0840 00) is available for the connection of system-external
call devices. In addition to triggering the "Call", "Alarm" or "WC Call" call type, this
interface provides the location light and reassurance light function, as well as a bed
identification function.