14. Servicing
Inspection measures
Call switch/WC with cancel switch
Inspection interval: 3 months
Is the housing undamaged and free of
contamination?
Can a call be triggered by pressing the red
call button?
Do the location and reassurance lights of
the red call switch work?
Does the switch emit an acoustic acknowl-
edgement of the call initiation?
Can a WC call be cancelled using the grey
cancel button?
Does the LED next to the cancel button
light up when a WC call that can be can-
celled with this cancel button is activated?
If the optional "Call tone" function is acti-
vated via configuration on this switch,
check additionally: Does the switch emit a
call tone after an overflow time during ac-
tivated staff presence if a call has been in-
itiated in another room of the ward?
Overflow time according to operating
mode and configuration in the SystemOr-
ganizer.
Call switch with privacy switch
Inspection interval: 3 months
Is the housing undamaged and free of
contamination?
Can a call be triggered by pressing the red
call button?
Do the location and reassurance lights of
the red call button work?
Does the switch emit an acoustic acknowl-
edgement of the call initiation?
80
Order no.
77 0217 00 A
77 0217 00 C
77 0217 00 F
77 0218 00 A
77 0218 00 C
77 0218 00 F
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)
77 0217 00 A
77 0218 00 A