14.7

Patient devices

Functions
Patient handset
Inspection interval: 3 months
 Are the housing, buttons, connection ca-
ble and plug undamaged and free of con-
tamination?
 Can a call be triggered by pressing the red
call button?
 Do the location and reassurance lights of
the red call button work?
 Is a speech connection to the patient
handset established when a call is an-
swered?
 Can both partners in the conversion hear
at an appropriate volume and understand
each other?
 Does the sound become louder when you
hang the patient handset into a patient
handset bracket (70 0800 00)?
 Is the plug connection stable when you
connect headphones to the headset jack
of the patient handset?
 Is the sound forwarded to connected
headphones?
Pear push switch incl call and light
switch
Inspection interval: 3 months
 Are the housing, buttons, connection ca-
ble and plug undamaged and free of con-
tamination?
 Can a call be triggered by pressing the red
call button?
 Do the location and reassurance lights of
the red call button work?
Pear push switch incl. 2 call switches
Inspection interval: 3 months
 Are the housing, buttons, connection ca-
ble and plug undamaged and free of con-
tamination?
 Can a call be triggered by pressing the red
call button? Check both buttons.
 Do the location and reassurance lights
work?
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)
Order no.
74 0747 00
70 0710 00
70 0710 01
70 0711 00
70 0711 01
14. Servicing
89