Connection socket combi
Inspection interval: 12 months
 Is the housing undamaged and free of
contamination?
 Do the plug sockets sit precisely and firm-
ly?
 Is the plug connection stable when a pa-
tient handset is connected to the large
plug socket?
 Does the patient handset work when it is
connected to the connection socket? Test
as described for the patient handset.
 Is the plug connection stable when a plug-
gable call device (e.g., pear push switch)
is connected to the small plug socket?
 Does the pluggable call device (e.g., pear
push switch) work when it is connected to
the connection socket? Test as described
for the pluggable call device.
Connection socket combi, bedhead
unit
Like connection socket combi.
Connection socket ComStation
Inspection interval: 12 months
 Is the housing undamaged and free of
contamination?
 Does the plug socket sit precisely and
firmly?
 Is the plug connection stable when the
corresponding device (ComStation
CT
ComStation
Flamenco, ComStation
Flamenco, ComTerminal Flamenco in
tabletop design or RoomTerminal Fla-
menco in tabletop design) is connected to
the connection socket?
 Are the screws of the connecting plug
screwed into the socket?
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)
Order no.
70 0424 00
70 0425 00
70 0424 50
70 0434 00
70 0435 00
70 0434 50
74 0452 30
BUS-C
,