14.4
Switches
Inspection measures
Call switch
Inspection interval: 3 months
Is the housing undamaged and free of
contamination?
Can a call be triggered by pressing the red
call button? Check both buttons.
Do the location and reassurance lights of
the call switch work?
Does the switch emit an acoustic acknowl-
edgement of the call initiation?
WC call switch
Like 77 0211 00 A, but a WC call is triggered.
Staff presence switch
Inspection interval: 12 months
Is the housing undamaged and free of
contamination?
Can presence 1 be activated and deacti-
vated using the green presence switch?
Does the reminder light work?
Can presence 2 be activated and deacti-
vated using the green presence switch?
Does the reminder light work?
If the optional "Call tone" function is acti-
vated via configuration on this switch,
check additionally: Does the switch emit a
call tone after an overflow time during ac-
tivated staff presence if a call has been in-
itiated in another room of the ward?
Overflow time according to operating
mode and configuration in the SystemOr-
ganizer.