14. Servicing
14.5
Room terminals
Functions
ComTerminal Flamenco
Inspection interval: 3 months
Is the housing undamaged and free of
contamination?
Can presence 1 be activated and deacti-
vated using the green presence key?
Does the reminder light work?
Can presence 2 be activated and deacti-
vated using the yellow presence key?
Does the reminder light work?
If presence is activated: Can a cardiac
alarm be triggered by pressing the blue
alarm key? Prerequisite: According to the
configuration, the alarm key is available
on the ComTerminal.
Do the location (in case of presence) and
reassurance lights of the blue alarm key
work?
Can a call be triggered by pressing the red
call button?
Do the location and reassurance lights of
the red call button work?
If presence is activated: Does the
ComTerminal emit a call tone after an
overflow time if a call has been trigged in
another room of the ward? Overflow time
according to operating mode and configu-
ration in the SystemOrganizer.
Are the call type and call location shown
on the display?
Can you answer the call by pressing the
corresponding key on the ComTerminal?
For calls with speech possibility: Is a
speech connection to the call location es-
tablished when a call is answered?
Can both conversation partners hear at an
appropriate volume and understand each
other?
For the "Call" call type: Can you cancel the
call remotely?
Continued on the next page.
82
Order no.
77 0510 00
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)