Tunstall FlamencoIP Technical Manual page 96

Hide thumbs Also See for FlamencoIP:
Table of Contents

Advertisement

14. Servicing
Functions
Radio smoke detector
to be assigned to radio receiver-T
Recommendation of the manufacturer of the
radio receiver T: Weekly inspection with test
call reception under observance of the dis-
play elements and call forwarding.
 Function test according to the provided
documentation of the manufacturer.
 If the radio receiver of the smoke detector
indicates a low battery, the radio smoke
detector must be replaced.
 Is a call triggered in the nurse call system
when the smoke detector indicates the
emission of smoke?
Large-surface radio pneumatic switch
to be assigned to radio receiver-T
Recommendation of the manufacturer of the
radio receiver T: Weekly inspection with test
call reception under observance of the dis-
play elements and call forwarding.
 Are the housing and switch undamaged
and free of contamination?
 Can a call be triggered by pressing the red
trigger button?
 If the radio receiver for the pneumatic
switch indicates a low battery, the battery
must be replaced.
 Check whether the pneumatic switch
works across the entire intended area of
application (range test).
94
Order no.
68005/96
75 0711 00
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)