Functions
Sensor mat
plug-in connection to connection socket
Inspection interval: 3 months
 Are the mat, connection cable and plug
undamaged and free of contamination?
 Is a call triggered when you step on or
press the mat?
 Has it been guaranteed that no objects
are located on the mat?
 Is the mat placed in a non-slipping man-
ner?
Breathing sensor set
plug-in connection to connection socket
Function test according to the provided docu-
mentation of the manufacturer.
At minimum, a weekly inspection with test call
triggering, an optical inspection of the con-
nection to the nurse call system and a man-
ual test of the holding force of the mounting
are suggested. After a patient is moved or the
system is moved to an environment with
other interfering noises, the settings and
placement must be performed again and
checked.