14.11
Interfaces
14.11.1
Interfaces in the group
Functions
OSYlink-Universal
Inspection interval: 3 months
Is the connected external device undam-
aged and free of contamination?
If a call device is connected: Is a call trig-
gered in the nurse call system if the call
device is triggered?
If present: Do the location and reassur-
ance lights of the call device work?
Can the call be cancelled?
14.11.2
Interfaces in the room
Functions
RAN interface
Inspection interval: 3 months
Is the connected external device undam-
aged and free of contamination?
If the external device is a call device: Is a
call triggered in the nurse call system if the
call device is triggered?
If present: Do the location and reassur-
ance lights of the call device work?
Can the call be cancelled?
If the external device is a presence detec-
tor: Can presence 1 be activated and de-
activated using the presence detector?
Does the reminder light work?
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)
Order no.
77 0803 00
Order no.
77 0840 00