14. Servicing
Functions
T
ComStation
Flamenco
Inspection interval: 3 months
Are the housing and connection cable un-
damaged and free of contamination?
Is the plug of the connection cable firmly
seated in the socket and secured with
both screws?
Is the display easily readable?
Is a fault in the nurse call system shown
on the display? If so, locate the fault and
remedy it.
Can presence 1 be activated and deacti-
vated using the green presence key?
Does the reminder light work?
Can presence 2 be activated and deacti-
vated using the yellow presence key?
Does the reminder light work?
If presence is activated: Can a cardiac
alarm be triggered by pressing the blue
alarm key?
Do the location (in case of presence) and
reassurance lights of the blue alarm key
work?
Can a call be triggered by pressing the red
call key?
Do the location and reassurance lights of
the red call key work?
If presence is activated: Does the Com-
T
Station
emit a call tone after an overflow
time if a call has been trigged in another
room of the ward? Overflow time accord-
ing to operating mode and configuration in
the SystemOrganizer.
Are the call type and call location shown
on the display?
Can you acknowledge the call by pressing
the corresponding key on the ComSta-
T
tion
?
100
Order no.
77 0606 20
IP
Flamenco
- Technical Manual - 00 8813 52, 03/18 (Rev. 1.0)