10.2.4
Call devices
Check whether calls from additional call devices (e.g. pear push switches and radio
triggers) are indicated as desired.
10.3

Checking the consoles

Perform this test for all installed consoles: Ward consoles, central consoles. For in-
formation on the correct operation and use of the consoles refer to the relevant op-
erating instructions or online help.
Two persons are required for executing these checks. One person will handle the
console, and the other person will proceed from one room to another.
Check for each room:
 Is the room number correctly displayed?
 Do all types of calls arrive at the console and are they signalled in the appropriate
manner (e.g. call type "Call", WC call, Diagnostic call, Emergency call 1, Emer-
gency call 2, WC emergency call, Cardiac alarm, Telephone call, Door call)?
 Are bed numbers (channel numbers) correctly presented?
 Can calls be answered at the console?
 Can a speech connection be established with the console via speech devices
(ComTerminal, patient handset)?
 Are presence states correctly displayed?
 Can announcements be broadcast from the console?
NOTE! If the wrong bed number is set on the connection socket bedhead unit, the
medical supply unit must be opened and the correct number must be set (see med-
ical supply unit documentation).
10.4

Checking functions between wards

Check:
 Can ward coupling be activated at the ward console?
 Correct function of group lamps.